Tesec NSG 4070 Immunity BCI Test System

Overview

  • Integrated EMC conducted RF immunity test system with built-in RF generator, power meter, and amplifier control
  • Designed for conducted immunity testing according to IEC / EN 61000-4-6
  • Supports standalone operation or remote control via PC interface
  • 5.7-inch TFT color display with intuitive menu-driven user interface
  • Supports external EUT monitoring and test automation

RF Generator Specifications

  • Frequency range: 9 kHz to 1 GHz
  • Frequency resolution: 1 Hz
  • Reference frequency: 10 MHz
  • RF output level range: –60 dBm to +10 dBm
  • Level resolution: 0.1 dB
  • Typical settling time: 10 ms

Modulation Capabilities

  • Amplitude modulation depth: 0 to 100%
  • Modulation frequency range: 1 Hz to 50 kHz
  • Pulse modulation rise and fall time: < 1 µs
  • Pulse duty cycle range: 10% to 90%
  • External modulation delay: < 1 µs / 180°

Power Meter Specifications

  • Frequency range: 9 kHz to 1 GHz
  • Linear measurement range (Channel 1): –35 dBm to +27 dBm
  • Linear measurement range (Channels 2 and 3): –45 dBm to +20 dBm
  • Maximum input (no damage): +28 dBm
  • Noise level: More than 5 dB below measurement range
  • Input connector: BNC, 50 Ω
  • Input return loss: > 20 dB below 500 MHz, > 17 dB from 500 MHz to 1 GHz

Power Amplifier Options

  • Available internal amplifier power levels: 20 W, 30 W, and 75 W (configuration dependent)
  • Nominal amplifier frequency range: 150 kHz to 230 MHz
  • Input and output impedance: 50 Ω
  • Gain flatness: ±3 dB
  • Maximum saturated output power: up to approximately 48.75 dBm (75 W version)

Standards and Applications

  • Conducted RF immunity testing according to IEC / EN 61000-4-6
  • Suitable for automotive BCI testing per ISO 11452-4
  • Applicable for EMC compliance, pre-compliance, and product development testing

Hire, Buy or More Information

Please fill in the below form and click “Submit”.

We are available on phone, Monday to Friday 9am GMT – 6pm GMT on +44 145 2245 709